IV Bags (filled)
Integrity flaws (welds, punctures)
Specifications and Customer requirements
100% inspection of pouches. The pouches are stored at a temperature between 2°C and 6°C, which causes condensation to form on their surface at ambient temperature. Detection of defects equivalent to a 50 µm hole. Quantity to be tested: 100 pieces per day.
Reference standards
ASTM F2095-07(2021) Standard Test Methods for Pressure Decay Leak Test for Flexible Packages With and Without Restraining Plates.
Solution
ASC 7400PP Leak Tester.
The ASC 7400PP allows for pouch testing, even when stored at low temperatures. By measuring the pressure inside the pouch (ASC Instrument patent), it eliminates false positives caused by condensation observed with other methods.
Operation
The operator places the pouch in the drawer and closes it. The cycle starts automatically and runs independently.
A vacuum is gradually applied inside the chamber, causing the pouch to inflate. It comes into contact with a sensor that measures the pressure inside the pouch. If the pressure remains stable during the testing phase, the pouch is leak-free, and the green light comes on. In case of a leak, the pressure in the pouch decreases, the measured pressure variation (delta P) exceeds the tolerance, and the red light indicates the defect.